Philips MP40/50 Manuel D’Utilisation

Page de 414
Viewing ST Maps
6 ECG, Arrhythmia, and ST Monitoring
125
Displaying an ST Reference Baseline
You can display an ST reference baseline for the current view, or for the trended view. The baseline is 
shown in yellow. However, if the ECG color is yellow, the baseline is shown in green. Use this baseline 
to detect ST changes. The baseline is derived automatically whenever the monitor relearns arrhythmia 
and also on user request.
Select 
Show Baseline
/
Hide Baseline
 to toggle between baseline display on and off.
Updating an ST Map Reference Baseline
To update the baseline,
In the 
Setup ST Analysis
 menu, select 
ST Baseline -> Update baseline
.
Changing the Scale of the ST Map
To change scale,
Select 
Size Up
 or 
Size Down
 to alter the size at which monitor displays the map.
Changing the Trending Interval
To determine how frequently the monitor displays a trended sample,
1
In Trend view, select 
Select Interval
.
2
Select the required interval from the menu. The interval ranges between 12 seconds to 30 minutes.
Printing an ST Map Report
To print the most recently viewed (current or trend) window, 
1
Select 
Main Setup
 - > 
Reports
.
2
Select 
ST Map
.
3
Press 
Print
.