Technical Alarm Messages (INOPs)
4 Patient Alarms and INOPs
75
tcpO
2
 (or tcpCO
or tcGas) 
EQUIP MALF 
Numeric is replaced by -?- 
INOP tone.
tcGas
There is a malfunction in the transducer or module. Connect 
another transducer. If this INOP persists, contact your service 
personnel.
tcpO
2
 (or tcpCO
or tcGas) 
NO TRANSDUC
 
Numeric is replaced by -?- 
INOP tone.
tcGas
No transducer is connected to the tcpO
2
/tcpCO
2
 module. 
Silencing the alarm switches off the measurement. 
tcpO
2
 (or tcpCO
or tcGas) 
STABILIZING 
Numeric is displayed with a ?
tcGas
The transducer has not yet reached the selected temperature 
and/or skin hyperemization is not yet finished. This INOP will 
disappear within three minutes.
tcpO
2
 (or tcpCO
or tcGas) 
UNPLUGGED 
Numeric is replaced by -?- 
INOP tone.
tcGas
The measurement is switched on but the module is unplugged. 
The measurement automatically disappears from the display. 
Silencing this INOP switches off the measurement.
<
Temp> CHK SOURCES
Numeric is replaced by -?- 
TEMP 
Difference 
Not all measurements or values required to perform the 
calculation are available. Check measurement sources.
<
Temp> CHK UNITS
Numeric is replaced by -?- 
TEMP 
Difference 
The monitor has detected a conflict in the units used for this 
calculation. Check the unit settings.
<Temp label> DEACTIVATED
INOP tone
TEMP
A Temp measurement label in the measurement server or 
extension has been deactivated, either by connecting a Pressure 
transducer in the shared Press/Temp socket, or by deactivating 
the label in the Measurement Selection window. 
The measurement automatically disappears from the display. 
To switch the measurement on again, either reconnect a Temp 
transducer or reactivate the measurement label in the 
Measurement Selection window. 
<Temp label> EQUIP MALF
Numeric is replaced by -?- 
INOP tone.
TEMP
Contact your service personnel.
The temperature hardware is faulty.
<Temp label> NO 
TRANSDUCER
Numeric is replaced by -?- 
INOP tone.
TEMP
Make sure the TEMP probe is connected to the MMS or 
module.
If you silence this INOP, the measurement will be switched off.
<Temp label> UNPLUGGED
INOP tone
TEMP
A Temp measurement label has been deactivated, either by 
unplugging a module, or by deactivating the label in the 
Measurement Selection window. 
The measurement automatically disappears from the display. 
To switch the measurement on again, either replug the module 
or reactivate the measurement label in the Measurement 
Selection window.
<Temp label> OVERRANGE
Numeric is replaced by -?- 
INOP tone.
TEMP
Try changing the application site of the transducer.
[The temperature is less than -1
°
C, or greater than 45
°
C.]
Tesoph
 INOPS
TEMP
See <Temp label> INOPs (under Temp).
TimeExpired:<
timer label
>
INOP tone
Monitor
The time has expired for the timer indicated in the INOP text. 
Clearing the timer clears the INOP. 
Tnaso
 INOPS
TEMP
See <Temp label> INOPs (under Temp).
Trect
 INOPS
TEMP
See <Temp label> INOPs (under Temp).
Tskin
 INOPS
TEMP
See <Temp label> INOPs (under Temp).
INOP Message, Indication
Source
What to do