Patient Alarm Messages
4 Patient Alarms and INOPs
55
**HR HIGH
ECG
The heart rate high alarm limit has been 
exceeded.
numeric flashes and high limit 
is highlighted, yellow alarm 
lamp, alarm tone. The sound 
switches off after 5 seconds if 
Arrhythmia is On.
**HR LOW
ECG
The heart rate has fallen below the low 
alarm limit.
numeric flashes and low limit 
is highlighted, yellow alarm 
lamp, alarm tone. The sound 
switches off after 5 seconds if 
Arrhythmia is On.
**imCO2 HIGH
CO
2
, Resp, 
AGM
The inspired minimum CO
2
 high alarm 
limit has been exceeded.
numeric flashes and high limit 
is highlighted, yellow alarm 
lamp, alarm tone.
**in <Agent label> 
HIGH
AGM
The inspired agent high alarm limit has 
been exceeded.
numeric flashes, high limit is 
highlighted, yellow alarm 
lamp, alarm tone.
**in <Agent label> 
LOW 
AGM
The inspired agent value has fallen 
below the AGT low alarm limit.
numeric flashes, low limit is 
highlighted, yellow alarm 
lamp, alarm tone.
**inN2O HIGH
N
2
O, AGM
The inspired N
2
O high alarm limit has 
been exceeded.
numeric flashes, high limit is 
highlighted, yellow alarm 
lamp, alarm tone.
**inO2 HIGH
O
2
, AGM
The inspired O
2
 high alarm limit has 
been exceeded.
numeric flashes, high limit is 
highlighted, yellow alarm 
lamp, alarm tone.
**inO2 LOW
O
2
, AGM
The inspired O
2
 value has fallen below 
the low alarm limit.
numeric flashes, low limit is 
highlighted, yellow alarm 
lamp, alarm tone. 
***inO2 LOW OXYGEN 
O
2
, AGM
The inspired O
2
 value has fallen below 
18 vol.%. 
numeric flashes, low limit is 
highlighted, red alarm lamp, 
alarm tone.
*/**IRREGULAR HR
 ECG/
Arrhythmia 
Consistently irregular heart rhythm.
numeric flashes, yellow alarm 
lamp, short yellow audible 
alarm.
*/**MISSED BEAT
ECG/
Arrhythmia
No beat detected for 1.75*R-R interval, 
or if HR>120bpm no beat detected for 
one second (non-paced patients only).
numeric flashes, yellow alarm 
lamp, short yellow audible 
alarm.
*/**MULTIFORM PVCs
ECG/
Arrhythmia
Two differently shaped Vs detected, 
each occurring at least twice within the 
last 300 beats and at least once within 
the last 60 beats.
numeric flashes, yellow alarm 
lamp, short yellow audible 
alarm.
**NBP HIGH
NBP
The measured NBP value is above the 
high alarm limit.
sd, or m after the label indicates 
whether the systolic, diastolic or mean 
pressure has crossed the limit.
numeric flashes and high limit 
is highlighted, yellow alarm 
lamp, alarm tone.
**NBP LOW
NBP
The measured NBP value is below the 
low alarm limit.
sd, or m after the label indicates 
whether the systolic, diastolic or mean 
pressure has crossed the limit.
numeric flashes and low limit 
is highlighted, yellow alarm 
lamp, alarm tone.
Alarm Message
From
Condition
Indication