Configuring the NOC Application
7. Set the parameters in the Compensations menu as desired.
For
 Transient Bubble Remediation
, see Section 2.3.4 for a discussion of this feature, and 
see Section 4.6 for additional setup instructions.
For 
Transient Mist Remediation
, see Section 2.3.6 for a discussion of this feature, and 
see Section 4.8 for additional setup instructions.
For 
Pressure Compensated Density
, see Section 2.3.2 for a discussion of this feature, 
and see Section 4.5 for additional setup instructions.
For 
Shrinkage Factors
, see Section 2.3.3 for a discussion of this feature. To “disable” 
shrinkage factors, set them to 1.0.
8. If desired, configure discrete outputs, milliamp outputs, or the frequency output to report NOC 
data. To do this, refer to the manual entitled Series 3000 MVD Transmitters and Controllers: 
Configuration and Use Manual 
and:
a.
Assign the desired process variable to the output. Available NOC process variables are 
listed in Table 4-2.
b.
Configure other output parameters as desired.
Table 4-2
Assigning NOC process variables to outputs
NOC process variable
Can be assigned to
Discrete outputs
Milliamp outputs
Frequency output
TBR event
(1)
(1) ON = TBR active; OFF = TBR inactive.
External water cut
Uncorrected oil volume rate
Uncorrected water cut
Uncorrected water volume rate
Back flow rate
Net oil flow rate
Net water cut
Gross volume rate
Net water flow rate
Average uncorrected oil rate
Average uncorrected water cut
Average uncorrected gross rate
Average uncorrected water rate
Average back flow rate
Average net oil flow rate
Average net water cut
Average gross volume rate
Average net water rate
Average gas volume flow rate