30
Micro Motion
®
Series 3000 MVD Transmitters
Configuring the NOC Application
4.5.2
Setting up pressure compensation for oil density and water density
To set up pressure compensation for oil density and water density:
1. See Figure 4-1. In the Well Data-Densities menu, set
Press Comp Oil Density
to the factor to
be used to compensate for the effect of pressure on oil density.
2. In the Compensations menu, set
Press Comp Density
to
Enable
.
4.5.3
Setting up pressure compensation for pressure effect
To set up pressure compensation for pressure effect:
1. See Figure 3-2. Select
Configuration>Inputs>External Inputs
, and:
a.
Select the
Polling Variable
that you have defined for pressure.
b.
Set
Pressure Compensation
to
Enable
.
c.
Set
Flow Factor
,
Density Factor
, and
Cal Pressure
to the appropriate values for your
sensor. See Section 2.3.2.
4.6
Setting up Transient Bubble Remediation
To set up TBR:
1. See Figure 4-1. In the Compensations menu, select
Transient Bubble Remd
, and:
a.
Set
Drive Gain Level
to the drive gain value (in percent) that represents a transient bubble
condition in this process stream. To determine the best value, you may find it useful to
observe drive gain values for this system using the View>Well Performance
Measurements>View Production Measurements menu.
observe drive gain values for this system using the View>Well Performance
Measurements>View Production Measurements menu.
b.
Set
Action Taken
to the desired action.
c.
Set
Time Period
to the number of seconds that the transmitter will look back to retrieve a
density value.
2. If desired, configure a discrete output to report TBR status. See Section 4.3, Step 8.
4.7
Setting up gas measurement
To set up and enable gas measurement:
1. Ensure that the external gas measurement device is correctly installed and tested, and
configured to report gas data in the units that will be used in the NOC application.
2. Ensure that the external gas measurement device has been wired to the frequency input of the
Series 3000 platform.
3. At the Series 3000 platform, configure the frequency input and perform a loop test, as
described in the manual entitled Series 3000 MVD Transmitters and Controllers:
Configuration and Use Manual.
Configuration and Use Manual.
4. See Figure 4-1. In the External Inputs menu:
a.
Set
Gas Meter
to
Enable
b.
Set
Gas Volume Units
to the units to be used for gas measurement.