Emerson MMI-20011276 Manuel D’Utilisation

Page de 78
Net Oil Computer Supplement
33
Configuring the NOC Application
NOC Ov
er
vie
w
NOC Configuration
Using the Displa
y
Be
fo
re
 Y
o
Begin
NOC Ov
er
vie
w
NOC Configuration
Using the Displa
y
Be
fo
re
 Y
o
Begin
NOC Ov
er
vie
w
NOC Configuration
Using the Displa
y
Be
fo
re
 Y
o
Begin
NOC Ov
er
vie
w
NOC Configuration
Using the Displa
y
Be
fo
re
 Y
o
Begin
In-line density determination for water
To perform a in-line density determination for water:
1. Ensure that the correct fluid (water) is flowing through the sensor. You may need to close the 
outlet valve from the separator and wait for the phases to settle, then open the outlet valve.
2. From the Calibration menu, select 
Density Determination
 and work through the screens until 
the Procedure Selection screen (see Figure 4-3) is displayed. In Well Test mode, you will be 
required to select the well for which the procedure will be performed.
3. From the Procedure Selection screen, select 
Water Density
.
4. From the next screen, select 
Measure and Save
.
5. The display now shows the current values for water density, water temperature, volume, 
flowrate, and pressure (if pressure compensation for water density is enabled), plus the 
configured Water Deviation and Water Duration Ave values (see Figure 4-4). Monitor the 
density and temperature values, watching for the density and temperature readings to stabilize.
Note: The Water Deviation and Water Duration Ave values shown here are read-only. If you need to 
change them, you must reconfigure the well data. See Section 4.3.
6. (Optional) Press 
RESET
 to reset the volume total to 0. This enables you to monitor the amount 
of fluid that remains in the separator (if the separator volume is known).
Figure 4-4
Density determination for water – Preparing to start
7. When the density and temperature readings have stabilized, press 
START
. The NOC 
application will now average the density and temperature of the process fluid until a good 
sample is achieved. 
A “good sample” means that, for the configured Water Duration Ave, no two consecutive 
density readings differed by more than the configured Water Deviation limit. If the water 
deviation limit is exceeded, averaging is restarted. 
Note: If the averaging procedure does not end automatically, you may need to verify your process 
fluid or increase the Water Deviation value.
Measure & Save
Actual Rate
xxxx.x bbl/day
Water Deviation
x.xxxx g/cm3
Water Duration Ave
xx sec
START
RESET
EXIT
Measure & Save
Actual Water Density
x.xxxx g/cm3
Actual Temperature
xxx.x deg F
Actual Pressure
xx.xxxx PSI
Volume
xxxx.x bbl/day
START
RESET
EXIT
Top of screen
Bottom of screen