36
Micro Motion
®
 Series 3000 MVD Transmitters
Configuring the NOC Application
9. The NOC application converts the observed water density to water density at reference 
temperature, and displays a screen similar to Figure 4-8. This screen also displays the stored 
value for water density at reference temperature, and the date and time at which this value was 
stored.
To replace the stored value with the new value, press 
SAVE
.
To discard the data from this procedure, press 
EXIT
. You will be returned to the Procedure 
Selection screen. From this point you can repeat the density determination procedure or 
press 
EXIT
 to move up the menu and return to other functions.
Figure 4-8
Density determination for water – Manual procedure results
4.9.2
Performing a density determination for oil
To perform a density determination for oil:
1. Ensure that the correct fluid (oil) is flowing through the sensor. You may need to drain water 
from the separator.
2. From the Calibration menu, select 
Density Determination
 and work through the screens until 
the Procedure Selection screen (see Figure 4-3) is displayed. In Well Test mode, you will be 
required to select the well for which the procedure will be performed.
3. From the Procedure Selection screen, select 
Oil Density
. The screen shown in Figure 4-9 i
displayed.
Manually Enter
Water Density @ Ref
x.xxxx g/cm3
Water Density at 
dd-mm-yy hh:mm
Current Dens @ Ref
x.xxxxx g/cm3
Current Dens Saved
dd-mm-yy hh:mm
SAVE
HELP
EXIT