Emerson 2200S Manuel D’Utilisation

Page de 158
108
Micro Motion
®
 Model 2200S Transmitters
Troubleshooting
A008
Density Outside 
Limits
The measured density has 
exceeded 0–10 g/cm
3
• If other alarms are present (typically, A003, A006, A102, or 
A105), resolve those alarm conditions first. If the A008 
alarm persists, continue with the suggestions here.
• Verify process. Check for air in the flow tubes, tubes not 
filled, foreign material in tubes, or coating in tubes.
• Check for slug flow. See Section 10.17.
• Check the sensor circuitry. See Section 10.24. 
• Verify calibration factors in transmitter configuration. See 
• Check the test points. See Section 10.23.
• If the problem persists, contact Micro Motion.
Density Overrange
A009
Transmitter 
Initializing/Warming 
Up
Transmitter in power-up 
mode.
• Allow the flowmeter to warm up (approximately 30 
seconds). The error should disappear once the flowmeter 
is ready for normal operation.
• If alarm does not clear, make sure that the sensor is 
completely full or completely empty. 
• Check the sensor circuitry. See Section 10.24.
Transmitter 
Initializing/Warming 
Up
A010
Calibration Failure
Mechanical zero: The 
resulting zero was greater 
than 3 
μs.
Temperature/Density Cals: 
many possible causes.
• If alarm appears during a transmitter zero, ensure that 
there is no flow through the sensor, then retry.
• Cycle power to the flowmeter, then retry.
• If appropriate, restore the factory zero to return the 
flowmeter to operation.
Calibration Failure
A011
Excess Calibration 
Correction, Zero 
too Low 
See A10
• Ensure that there is no flow through the sensor, then retry.
• Cycle power to the flowmeter, then retry.
• If appropriate, restore the factory zero to return the 
flowmeter to operation.
Zero Too Low
A012
Excess Calibration 
Correction, Zero 
too High
See A10
• Ensure that there is no flow through the sensor, then retry.
• Cycle power to the flowmeter, then retry.
• If appropriate, restore the factory zero to return the 
flowmeter to operation.
Zero Too High
A013
Process too Noisy 
to Perform Auto 
Zero
See A10.
• Remove or reduce sources of electromechanical noise, 
then retry. Sources of noise include:
- Mechanical pumps
- Pipe stress at sensor
- Electrical interference
- Vibration effects from nearby machinery
• Cycle power to the flowmeter, then retry.
• If appropriate, restore the factory zero to return the 
flowmeter to operation.
Zero Too Noisy
A014
Transmitter Failed
Many possible causes.
• Cycle power to the flowmeter.
• The transmitter might need service. Contact Micro Motion. 
Transmitter Failed
A016
Line RTD 
Temperature 
Out-Of-Range
The value computed for 
the resistance of the Line 
RTD is outside limits
• Check the sensor RTD circuitry. See Section 10.24.
• Verify that process temperature is within range of sensor 
and transmitter.
• If the problem persists, contact Micro Motion. 
Line RTD 
Temperature 
Out-of-Range
A020
Calibration Factors 
Unentered
The flow calibration factor 
and/or K1 has not been 
entered since the last 
master reset.
• Check the characterization. Specifically, verify the FCF 
and K1 values. See Section 4.2.
• If the problem persists, contact Micro Motion. 
Calibration Factors 
Unentered 
(FlowCal)
Table 10-3
Status alarms and remedies continued
Alarm 
code
Communicator 
Cause
Suggested remedy
ProLink II