Erratic non-zero flow rate under 
no-flow conditions
Leaking valve or seal
• Check pipeline.
Slug flow
Plugged flow tube
• Check drive gain and tube frequency. 
Purge the flow tubes.
Incorrect sensor orientation
• Sensor orientation must be 
appropriate to process fluid. See the 
installation manual for your sensor.
Wiring problem
• Check the sensor circuitry. See 
Vibration in pipeline at rate close to 
sensor tube frequency
• Check environment and remove 
source of vibration.
Damping value too low
• Check configuration. See 
Mounting stress on sensor
• Check sensor mounting. Ensure:
- Sensor is not being used to support 
pipe.
- Sensor is not being used to correct 
pipe misalignment.
- Sensor is not too heavy for pipe.
Sensor cross-talk
• Check environment for sensor with 
similar (±0.5 Hz) tube frequency.
Erratic non-zero flow rate when flow 
is steady
Slug flow
Damping value too low
• Check configuration. See 
Plugged flow tube
• Check drive gain and tube frequency. 
Purge the flow tubes.
Excessive or erratic drive gain
• See Section 10.23.2
Output wiring problem
• Verify wiring between transmitter and 
receiving device. See the installation 
manual for your transmitter.
Problem with receiving device
• Test with another receiving device.
Wiring problem
• Check the sensor circuitry. See 
Inaccurate flow rate or batch total
Bad flow calibration factor
• Verify characterization. See 
Inappropriate measurement unit
• Check configuration. See 
Bad sensor zero
• Rezero the flowmeter. See 
Bad density calibration factors
• Verify characterization. See 
Bad flowmeter grounding
Slug flow
Problem with receiving device
• See Section 10.16.
Wiring problem
• Check the sensor circuitry. See 
Table 10-4
Process variables problems and remedies continued
Symptom
Cause
Suggested remedy