Emerson 2200S Manuel D’Utilisation

Page de 158
116
Micro Motion
®
 Model 2200S Transmitters
Troubleshooting
10.22
Checking the calibration
Improper calibration can cause the transmitter to send unexpected output values. If the transmitter 
appears to be operating correctly but sends inaccurate output values, an improper calibration may be 
the cause.
Micro Motion calibrates every transmitter at the factory. Therefore, you should suspect improper 
calibration only if the transmitter has been calibrated after it was shipped from the factory. Before 
performing a calibration, consider meter validation or meter verification and select the appropriate 
procedure (see Section 9.2). Contact Micro Motion customer service for assistance.
10.23
Checking the test points
Some status alarms that indicate a sensor failure or overrange condition can be caused by problems 
other than a failed sensor. You can diagnose sensor failure or overrange status alarms by checking the 
flowmeter test points. The test points include left and right pickoff voltages, drive gain, and tube 
frequency. These values describe the current operation of the sensor.
10.23.1
Obtaining the test points
You can obtain the test points with a Communicator or ProLink II.
With a Communicator
To obtain the test points with a Communicator:
1. Select 
Diag/Service
.
2. Select 
Test Points
.
3. Record the values displayed for 
Drive
LPO
RPO
, and 
Tube
With ProLink II
To obtain the test points with ProLink II:
1. Select 
Diagnostic Information
 from the 
ProLink
 menu.
2. Record the values displayed for 
Tube Frequency
Left Pickoff
Right Pickoff
, and 
Drive 
Gain
.
10.23.2
Drive gain problems
Problems with drive gain can appear in several different forms:
Saturated or excessive (near 100%) drive gain 
Erratic drive gain (e.g., rapid shifting from positive to negative)
Negative drive gain
See Table 10-5 for a list of possible problems and remedies.