Chapter 6
Integrating the Meter with the Control System
6.1
Overview
This chapter discusses the following topics and tasks:
Configuring the mA output – see Section 6.2
Configuring digital communications – see Section 6.3
Before beginning configuration, make an administrative connection to the transmitter and ensure that 
you are complying with all applicable safety requirements.
6.2
Configuring the mA output
The mA output is used to report a process variable. The mA output parameters control how the 
process variable is reported.
Table 6-1 lists the parameters that must be set for the mA output, and shows the names used for each 
parameter by the display, the Communicator, and ProLink II.
Note: If you use the display, you can configure only the process variable, LRV, and URV.
For details on mA output parameters, see Sections 6.2.1 through 6.2.4.
Table 6-1
mA output configuration parameters
Parameter
Parameter name
Display
Communicator
ProLink II
Process variable
SRC
PV
Primary variable
Range
12 mA
PV LRV
Lower range value
20 mA
PV URV
Upper range value
AO cutoff
Not accessible
PV AO cutoff
AO cutoff
Added damping
Not accessible
PV AO added damp
AO added damp
Fault action
Not accessible
AO1 fault indicator
AO fault action
Fault value
Not accessible
mA1 fault value
AO fault level