Emerson 2200S Manuel D’Utilisation

Page de 158
78
Micro Motion
®
 Model 2200S Transmitters
Testing and Moving to Production
Figure 7-6
ProLink II – Loop test procedure
7.4
Trimming the milliamp output
Trimming the mA output creates a common measurement range between the transmitter and the device 
that receives the mA output. For example, a transmitter might send a 12 mA signal that the receiving 
device reports incorrectly as 12.2 mA. If the transmitter output is trimmed correctly, it will send a 
signal appropriately compensated to ensure that the receiving device actually indicates a 12 mA 
signal.
Note: If a barrier or the Micro Motion adapter-barrier is installed, the mA output trim procedure can 
be extended to include additional trimming and calibration. See Section 7.4.2.
You must trim the mA output at both the 12 mA and 20 mA points to ensure appropriate 
compensation across the entire output range.
7.4.1
Basic mA output trim
To trim the output:
Using the Communicator, see Figure 7-7.
Using ProLink II, see Figure 7-8.
Additionally, if you are using a Communicator, you can perform a scaled AO trim. The scaled AO 
trim is used when the reference meter’s low and high values are not 4 and 20 mA. To perform a scaled 
AO trim, see Figure 7-9. 
Fix Milliamp
Enter mA value
Read output at receiving 
device
Correct?
Check output wiring
Troubleshoot receiving device
Yes
No
     ProLink Menu >
          Test 
Fix mA
Unfix
Set mA measurement point