Rosemount 4500
2-8
Grounding 
Transmitter Case
Always ground the transmitter case in accordance with national and local 
electrical codes. The most effective transmitter case grounding method is a 
direct connection to earth ground with minimal impedance. Methods for 
grounding the transmitter case include:
Internal Ground Connection: The Internal Ground Connection screw 
is inside the electronics housing. The screw is identified by a ground 
symbol (
), and is standard on all 4500 transmitters.
INSTALLING THE 
LCD METER
Transmitters ordered with the LCD display will be shipped with the display 
installed. Installing the display on an existing 4500 transmitter requires a small 
instrument screwdriver and the display kit.
The optional LCD display can be rotated in 90-degree increments.
Use the following procedure to install the LCD display:
1. IF the transmitter is installed in a loop, THEN secure the loop and 
disconnect power.
2. Remove the transmitter cover.
3. Engage the four-pin connector from the LCD display into the four-pin 
receptacle. Tighten screws to secure the LCD.
Note the following LCD temperature limits:
Ambient
32 to 140 °F (0 to 60 °C)
Storage
–22 to 185 °F (–30 to 85 °C)
Figure 2-6. Optional LCD 
Display