Rosemount Analytical Inc.    A Division of Emerson Process Management                                                                         Operation     3-5
Model 755A
2.  A zero-suppressed range uses a
blend consisting of a suitable
percentage of oxygen contained in a
background gas, typically nitrogen.
The upscale standard gas is required to
establish a calibration point at or near the
upper range limit.  For example, if this
range limit is 21% (or somewhat greater
than 21%), the usual upscale standard
gas is dry air (20.93% oxygen).
Typical examples of standard gases for
recorder oxygen ranges are shown in
Table 3-2, page 3-6.
A.  TYPICAL ZERO-BASED RANGES
0 to 1
Nitrogen
0.9% O2
Balance N2
0 to 10
Nitrogen
9% O2
Balance N2
0 to 100
Nitrogen
100% O2
B.  TYPICAL ZERO-SUPPRESSED RANGES
RANGE
% O2
RECOMMENDED STANDARD GAS
RECOMMENDED UPSCALE
STANDARD GAS
20 to 21
20.2% 
±0.2% O2
Air (20.93% O2)
11 to 21
11.2% 
±0.2% O2 Balance N2
Air (20.93% O2)
99 to 100
High Purity O2 with 0.8% 
±0.05% High Purity N2
High Purity O2
90 to 100
91% 
±0.5% O2 Balance N2
High Purity O2
Table 3-1. Standard Gases Recommended for Calibration of Various Oxygen Ranges on Analog Output