Invacare IRC5PO2V Manuel D’Utilisation

Page de 32
SECTION 4—TYPICAL PRODUCT PARAMETERS
12
SECTION 4—TYPICAL PRODUCT 
PARAMETERS
Electrical Requirements:
120 VAC ± 10% (132 VAC/108 VAC), 60 Hz
Rated Current Input:
3 A 
Sound Level:
43 ± 2 dBA Average for Perfecto
2
 Models and 
Perfecto
2
V Model
40 ± 2 dBA Average for Perfecto
2
W Models
Altitude:
All Perfecto
2
 Models: Up to 8,000 ft (2438 
meters) above sea level without degredation of 
concentration levels. From 8,000 ft to 13,129 
(4000 meters) below 90% efficiency.
All Perfecto
2
 Models while filling a 
HomeFill System: Up to 6,000 ft (1828 meters) 
above sea level without degradation of 
concentration levels. From 6,000 ft (1828 meters) 
to 13,129 ft (4000 meters) below 90% efficiency.
*Oxygen Output
Concentration Levels:
87% to 95.6% for flows ranging at 0.5 to 5 L/min
Maximum Outlet Pres-
sure:
5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)
Flow Range:
0.5 to 5 L/min 
For flowrates less than 1 L/min, we recommend 
the use of the Invacare Pediatric Flowmeter 
IRCPF16 Accessory.
Low Flow Alarm:
0 L/min to 0.5 L/min 
Rapid Audible Alarm Beeping (No Accessories 
Connected).
Average Power 
Consumption:
Perfecto
2
 and Perfecto
2
W Models: 280 W 
(Operating @ 3 L/min)
Perfecto
2
V Model: 325 W
Pressure Relief Mecha-
nism Operational at:
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)