2.2.2 Installing the analyzer
When used in Model DO-03/04, the 54eA analyzer is intend-
ed to be mounted on the basin handrail. See Figures 2-1
and 2-2.
ed to be mounted on the basin handrail. See Figures 2-1
and 2-2.
2.2.3 Installing the air compressor enclosure
The air compressor is in a NEMA 4X fiberglass enclosure.
Figure 2-3 shows how to attach the enclosure to a handrail.
Figure 2-3 shows how to attach the enclosure to a handrail.
1. Locate the enclosure within 25 ft (7.6 m) of the sensor.
2. Be sure the ambient temperature is between 5 and
122°F (-15 to 50°C). Do not expose the enclosure to
direct sunlight. The operating temperature for the air
compressor might be exceeded.
direct sunlight. The operating temperature for the air
compressor might be exceeded.
2.2.4 Installing the handrail mounting assembly
The handrail mounting assembly (PN 24048-00) is shipped completely disassembled. Assemble and attach the
mounting assembly to the handrail. Refer to Figure 2-4. The spring pin shipped with the handrail mounting
assembly is not needed. Do not attach the boom clamp to the clevis yet. Attach the maintenance bracket (PN
24044-00) to the handrail. See Figure 2-1.
mounting assembly to the handrail. Refer to Figure 2-4. The spring pin shipped with the handrail mounting
assembly is not needed. Do not attach the boom clamp to the clevis yet. Attach the maintenance bracket (PN
24044-00) to the handrail. See Figure 2-1.
2.2.5 Assembling the pipe and pipe clamp
Slide the PVC pipe boom through the handrail mounting pipe clamp. Tighten the bolt to keep the clamp from
sliding off. Use 1-1/2 inch schedule 80 PVC piping. Distance between the clamp and the end of the pipe should
not exceed 12 feet (3.6 m). See Figures 2-1 and 2-4.
sliding off. Use 1-1/2 inch schedule 80 PVC piping. Distance between the clamp and the end of the pipe should
not exceed 12 feet (3.6 m). See Figures 2-1 and 2-4.
8
FIGURE 2-2. Pipe Mounting
MODEL DO-03/04
SECTION 2.0
INSTALLATION
FIGURE 2-1. Suggested arrangement of handrail mounting assembly, maintenance clamp,
air compressor enclosure, and 54eA analyzer.