Emerson DO-03 Manuel D’Utilisation

Page de 70
MODEL DO-03/04
SECTION 5.0
SOFTWARE CONFIGURATION
5.2  RANGING THE OUTPUTS
2.  Press any key to enter the main menu. Move the cursor to "Program"
and press Enter (F4).
3.  Move the cursor to "Output setpoints" and press Enter (F4).
4.  Move the cursor to the desired output and press Enter (F4).
5.  This screen confirms that changes to output 1 are going to be made.
Press Cont (F3) to continue. Otherwise, press Abort (F1).
6.  This screen shows the present settings for Output 1. If the output
was configured to be 0-20 mA, the first line will show "0mA" instead
of "4mA". The live current output is shown on the third line. 
Move the cursor to the desired line and press Edit (F4). Use the
arrow keys to change the setpoint. Press Save (F4) to store the new
value.
Press Exit (F1) to return to the screen in step 4. Choose the other
output and continue.
Output 1 setpoints
Output 2 setpoints
Exit
Enter
Alarm setpoints
Output setpoints
Simulated test
Exit
Enter
CAUTION: Current
Output 1 will be 
affected.
Abort
Cont
1.  Ranging the outputs means assigning values to the low (0 or 4 mA) and high (20 mA) outputs. Before rang-
ing the outputs, be sure the outputs are properly configured. See Section 5.5. 
4 mA
:  0.00 ppm
20 mA:  20.00 ppm 
Output 1:  12.00 mA
Exit
Edit
25