Invacare IRC5PAW Manuel D’Utilisation
SECTION 4—TECHNICAL DESCRIPTION
Perfecto
2
™Series
12
SECTION 4—TECHNICAL
DESCRIPTION
DESCRIPTION
The Invacare Perfecto
2
concentrator is used by patients with
respiratory disorders who require supplemental oxygen. The
device is not intended to sustain or support life.
device is not intended to sustain or support life.
The oxygen concentration level of the output gas ranges from
87% to 95.6%. The oxygen is delivered to the patient through
the use of a nasal cannula.
87% to 95.6%. The oxygen is delivered to the patient through
the use of a nasal cannula.
The Invacare Perfecto
2
concentrator uses a molecular sieve
and pressure swing adsorption methodology to produce the
oxygen gas output. Ambient air enters the device, is filtered
and then compressed. This compressed air is then directed
toward one of two nitrogen adsorbing sieve beds.
Concentrated oxygen exits the opposite end of the active sieve
bed and is directed into an oxygen reservoir where it is
delivered to the patient.
oxygen gas output. Ambient air enters the device, is filtered
and then compressed. This compressed air is then directed
toward one of two nitrogen adsorbing sieve beds.
Concentrated oxygen exits the opposite end of the active sieve
bed and is directed into an oxygen reservoir where it is
delivered to the patient.
The Invacare Perfecto
2
concentrator is capable of operation by
the patient in a home environment or in an institutional
environment. Device operates at a nominal 230 VAC/50 Hertz
supply.
environment. Device operates at a nominal 230 VAC/50 Hertz
supply.
Service information will be available upon request to
qualified technical personnel only.
qualified technical personnel only.