Hazardous Area Oxymitter DR
 
APPENDIX A
UPGRADING HAZARDOUS AREA OXYMITTER DR TO
FULL HAZARDOUS AREA OXYMITTER
A-1 UPGRADE 
PROCEDURE
Perform the following procedure to upgrade the
Hazardous Area Oxymitter DR to a full Hazard-
ous Area Oxymitter.
a.  Remove power from the Oxymitter DR.
b.  Remove the left and right covers from the
Oxymitter termination housing.
c.  Remove and discard the two cable assem-
blies (1 and 2, Figure A-1) from the right
side of the termination housing.
d.  Place the new Oxymitter electronic assem-
bly (3) near the right side of the termination
housing.
e.  Plug the white connector with the two black
wires into the white socket on the bottom
power supply card.
f.  Insert the electronics assembly into the ter-
mination housing. Ensure the black 4-wire
connector remains outside the housing and
in the slot provided in the top card of the
electronics assembly. The electronics as-
sembly should seat on the bulkhead pins
easily. Do not force the assembly into place.
g.  Plug the black 4-wire connector into the
black socket on the microprocessor card.
h.  Tighten three screws (4) securing the elec-
tronics assembly into the termination hous-
ing.
2
36210017
1
3
DIA
DIA
GN
OS
TIC
GNOSTIC
ALA
RM
S
ALARMS
TE
ST
TEST
PO
INT
S
POINTS
HE
A
HEA
TE
R
TER
T/C
T/C
HE
A
HEA
TE
R
TER
02 C
ELL
02 CELL
CA
LIB
RA
CALIBRA
TIO
N
TION
CA
LIB
RA
CALIBRA
TIO
N R
EC
OM
ME
ND
ED
TION RECOMMENDED
02 C
ELL
 m
V +
02 CELL mV +
02 C
ELL
 m
v
02 CELL mv
-
HE
A
HEA
TE
R
TER
T/C
 +
T/C +
HE
A
HEA
TE
R
TER
T/C
 -
T/C -
INC
INC
INC
INC
DE
C
DEC
DE
C
DEC
HIG
H
HIGH
GA
S
GAS
LO
W
GA
S
GAS
CA
L
CAL
TE
ST G
AS
 +
TEST GAS +
PR
PR
OC
ES
S -
OCESS -
% 0
2
% 02
4
1. Signal connecting cable
2. Heater connecting cable
3. Electronic assembly
4. Screw
Figure A-1.  Component Replacement
A