Description and Specifications            1-1 
Model NGA2000 TO2 
SECTION 1 
DESCRIPTION AND SPECIFICATIONS 
 
 
 
 
 
 
1-1 OVERVIEW 
This manual describes the Trace Oxygen (TO2) 
Analyzer Module of Rosemount Analytical's 
NGA2000 Series of gas analysis components. 
The TO2 Analyzer Module is designed to con-
tinuously determine the concentration of trace 
oxygen in a flowing gaseous mixture. The con-
centration is expressed in parts-per-million. 
The TO2 Analyzer Module is configured as a 
shelf-mount module, designed to be installed 
external from the platform on an associated 
shelf capable of holding two modules side-by-
side, with gas connections made from the rear. 
All electronics relative to sample detection and 
conditioning are included in this module. 
1-2 TYPICAL APPLICATIONS 
The TO2 Analyzer Module has specific applica-
tions in the following areas: 
•  Trace oxygen in product nitrogen and argon 
streams from air separation plants 
•  Trace oxygen in inerting atmospheres for 
heat treat furnaces 
•  Trace oxygen in glove-box applications 
1-3  THEORY OF TECHNOLOGY 
The TO2 Analyzer Module uses the coulometric 
principle of oxygen detection. This technology is 
based on the fact that oxygen in the sample is 
reduced by an electrochemical reaction.  This 
reduction occurs at the cathode and results in 
the generation of hydroxyl ions.   These hy-
droxyl ions migrate to the anode where they are 
oxidized to reform oxygen.  The oxidation reac-
tion generates four electrons which in turn mi-
grate to the anode to participate in the reduction 
reaction: 
(Cathode Reaction) 
O
2
 + 2 H
2
O + 4 e
-
 
→  4 OH
-
 
(Anode Reaction) 
4 OH
-
 
→ O2 + 2 H
2
O + 4 e
-
 
A polarizing voltage of approximately 1.3 VDC  
is applied between the anode and cathode to 
drive the oxidation and reduction reactions.  The 
resulting current flow produced by the flow of 
electrons is directly proportional to the oxygen 
content in the sample gas. 
1-4 FEATURES 
Among the features included in the TO2 Ana-
lyzer Module are:  
•  Quick start feature 
•  Electrolyte level alarm  
•  High oxygen protection circuit with alarm 
•  Sample flow indication.