Emerson NGA2000 TO2 Manuel D’Utilisation

Page de 48
Instruction Manual 
748374-F 
March 2003 
1-4               Description and Specifications  
Rosemount Analytical Inc.    A Division of Emerson Process Management 
Model NGA2000 TO2
1-5 SPECIFICATIONS 
a. General 
Measurement Species................... Trace Oxygen 
Ranges .......................................... 0 to 100 ppm  (output scalable down to 0-2 ppm fullscale) 
Accuracy........................................
±3% of reading or ±0.02% of range (except for ranges ≤ 100 ppm: 
±3% of reading or ±0.05% of range) 
Sensitivity ...................................... <10 ppb Oxygen 
Noise ............................................. 1% of fullscale, peak to peak 
Linearity .........................................
±1% of fullscale 
Response Time ............................. Typically 90% in less than 20 seconds 
Zero Drift........................................
≤±1% of fullscale/24 hours at constant temperature 
Span Drift.......................................
≤±1% of fullscale/24 hours at constant temperature 
Effect of Temperature.................... 0.32% of reading per 
°F from 70°F 
(0.58% of reading per 
°C from 21°C) 
Effect of Flow.................................
≤2% of reading for a flow change of ±250 cc/min (0.5 SCFH) 
Operating Temperature ................. 32
°F to 113°F (0°C to 45°C) 
Power Requirements..................... +24 VDC 
±5%, 10 W max. 
Ripple and Noise: <100 mV peak to peak 
Line and Load Regulations: <
±1% 
 
b. Sample 
Sample .......................................... Non-flammable (below 100% of the LEL) 
Flow Rate ...................................... 0.5 to 1.5 L/min. 
Supply Pressure ............................ 1027 to 1082 hPa - absolute (0.2 to 1.0 psig) 
Temperature .................................. 32
°F to 113°F (0°C to 45°C) 
Particulates.................................... filtered to <0.1 mg/L; non-condensing at ambient temperature 
Sample Humidity ........................... non-condensing at ambient temperatures 
 
c. Physical 
Materials in contact with sample ... Stainless steel, Teflon, Delrin, neoprene 
Dimensions.................................... See Figure 2-2. Trace Oxygen Analyzer Outline and Mounting 
Dimensions on page 2-3 
Weight ........................................... 6.8 kg (15 lbs.) 
Mounting........................................ Horizontal, external to Platform or custom installed in a panel 
Case Classification ........................ General Purpose for installation in weather protected area 
Max. Separation from Platform...... 1600 m (1 mile) 
 
d. Gas 
Connections 
Sample In ...................................... 1/4 inch O.D. tube fitting 
Sample Out.................................... 1/4 inch O.D. tube fitting 
 
See the Preface Section of the Platform manual for specifications regarding Platform related components.