Instruction Manual 
748374-F 
March 2003 
2-4               Installation  
Rosemount Analytical Inc.    A Division of Emerson Process Management 
Model NGA2000 TO2
Sample
Inlet
1/4" Tube
Sample
Exhaust
1/4" Tube
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 2-3.  Trace Oxygen Analyzer Back Panel Connections 
 
 
 
 
CAUTION 
GAS OVERPRESSURE 
At no time should sample, zero or span 
gas inlet pressure exceed 69 hPa - gauge 
(1.0 psig). Damage to the Sensor may 
occur if this pressure level is exceeded. 
 
CAUTION 
SAMPLE FLOW 
Do not test the sample pressure by 
blocking the exhaust.  When the pres-
sure is released the sudden surge of 
flow will spin the internal flowmeter off 
its bearings and destroy it. 
 
c. LEAK 
TEST 
The TO2 Analyzer Module is completely 
tested at the factory for gas leakage. The 
user is responsible for testing for leakage 
only at the inlet and outlet fittings on the 
rear panel.  Caution: Do not expose the 
Sensor to pressure in excess of 1.0 psig 
as this may cause damage.