3-2          Startup and Operation 
Rosemount Analytical Inc.    A Division of Emerson Process Management 
Model NGA2000 TO2
From the Run Mode display, press the 
MENUS softkey to gain access to the Main 
Menu. 
The Basic controls menu is as follows: 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 3-3.  Basic Controls Menu 
 
This menu allows the user to view the cur-
rent range’s upper limit, Sleep mode, and 
quick start status.  It also allows quick start 
initiation or exit from sleep mode. 
In the figure above, the words in italics are 
the names of the network variables whose 
values are in fact shown on the screen. 
The Expert controls menu is as follows: 
 
 
 
 
 
 
Figure 3-4.  Expert Controls Menu 
 
This menu shows the current range number 
and range limits. 
The analyzer range settings may be config-
ured through the Analyzer Module set up 
menu, under Expert controls and set up
 
 
 
 
 
 
 
 
 
 
 
 
Figure 3-5.  Analyzer Module Set-up Menu 
 
c. Help 
Displays 
The Help structure is intended to be an on-
line "tutorial," context-sensitive and topic-
interconnected, so that the user can practi-
cally operate NGA2000 without the need of 
an instruction manual. 
A typical help menu: 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 3-6.  Typical Help Menu (shown is 
Zero/Span Calibration Help) 
 
This is the help screen for the calibration 
menus. 
 
 
Basic Controls
HOME
ESCAPE
INFO
Measurement range number:
Range upper limit:
Range and functional control:
Sleep Mode:
Exit Sleep Mode!
Initiate Quick Start!
Quick Start:
CURRENTRNGHI
CONTROL
 
 
Expert controls 
Measurement range number: 
Range lower limit: 
Range upper limit: 
 
Range and functional control: 
CURRENTRNGHI
CONTROL
HOME 
ESCAPE 
INFO 
Analyzer module set up
Range settings . . .
Displayed parameters . . .
Analyzer tag:
TAG
HOME
ESCAPE
INFO
 
 
Zero/Span calibration help 
This allows manual control of the zero and span.  Flow zero 
gas, and make sure the zero gas value is right; 
Press the zero key to make the analyzer zero. 
Or select Factors and individually adjust the reading on each 
range.  In this way you can make the analyzer read what you 
want. 
Do the same with span gas.  If the zero was not a real zero, 
the span action will change the zero reading> 
Set cal gas values in the Calibration gas list screen. 