Using AMS Device Manager to configure the meter
Section 4: Configuration
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014
3-9000-759 Rev D
6.
 Click the 
Frequency/Digital
 Outputs tab. Follow the configuration instructions in the
). (Refresh Note: If changes are made to 
any Source variable on this page, apply the changes and navigate to the Guided Setup 
page. Navigate back to the Manual Setup for the changes to be reflected in other 
Manual Setup pages). Click 
Apply
, after you enter the data to write the parameters to 
the meter. 
7.
Click the 
Frequency and Digital Output 1 
tab. Follow the configuration instructions in the 
b. Click 
Apply
, after you enter the data to 
write the parameters to the meter. 
8.
Click the 
Frequency and Digital Output 2 
tab. Follow the instructions in the 
 to configure the Frequency and Digital Output 2 
parameters. Click 
Apply
, after you enter the data to write the parameters to the meter. 
9.
Click the 
Temperature 
tab. Configure the input parameters including: Source (Live 
Analog or Fixed), Min and Max input limits corresponding to 4 mA and 20 mA 
respectively and the Low and High alarm limits. Click 
Apply
, after you enter the data to 
write the parameters to the meter. 
10.
Click the 
Pressure
 tab. Configure the input parameters including: Source (Live Analog or 
Fixed), Min and Max input limits corresponding to 4 mA and 20 mA respectively and the 
Low and High alarm limits. Select either 
Gage
 or 
Absolute
 for the type of pressure 
reading desired. If a live pressure transmitter is connected, select the type of reading 
the transmitter outputs. If Absolute is selected, you must also enter the Atmospheric 
pressure. Click 
Apply
, after you enter the data to write the parameters to the meter. 
11.
Click the 
Digital Input
 tab. The default Digital Input 1 polarity is set to 
Normal
 for general 
purpose or set to 
Inverted
 when used for calibration. Click 
Apply
, after you choose the 
calibration data to write the parameters to the meter. 
a.
Calibration Polarity configuration parameter selections are:
Digital Input 1 Calibrate Active High
Digital Input 1 Calibrate Active Low