Emerson 3414 Manuel D’Utilisation

Page de 144
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
Section 4: Configuration
3-9000-759 Rev D
June 2014
Using a Field Communicator to configure the meter
 97
5.
Wire Analog Input 1 (AI1) and Analog Output 1 (AO1) as shown in 
 and 
Appendix A
, drawing DMC-005324. 
Figure 4-20  Field Communicator wiring diagram for the 3410 Series electronics 
  
6.
Use the leads provided with the Field Communicator to connect to your device.
7.
Press and hold the 
Power
 button on the Field Communicator until the green light blinks.
8.
Use the touch screen on the Field Communicator, the keypad or use the stylus to 
navigate through the device menus.
9.
Refer to the Menu tree in Section D.1.1 of the Daniel HART Field Device Specification 
manual (P/N 3-9000-762) for the device fast key sequences. Included in the menu tree 
are:
Diagram Page 1 - 3410 Series Root Menu; Overview, Configure>Manual Setup 
Diagram Page 2 - Configure|Manual Setup (continued) and Alerts Setup
Diagram Page 3 - Service Tools|Alerts and Variables 
Diagram Page 4 - Service Tools|Variables (continued), Service Tools|Trends, and 
Service Tools|Maintenance
10.
If you encounter problems, refer to the contact information on the back cover of this 
manual or the contacts included in the Field Communicator User’s Manual.