Section 1: Introduction
3-9000-759 Rev D
June 2014
Safety
 19
Daniel 3410 Series Liquid Ultrasonic Meters are INMETRO certified. Refer to the 3410 Series 
Liquid Ultrasonic Flow Meter Tag, INMETRO Certification drawing DMC - 006224.
Certificate number: NCC 11.0163 X
Marking: --Ex d ia IIB T4 Gb IP66 W
Electrical parameters: Refer to 
Special conditions for safe use
Explosion proof joint dimensions are compliant with the Brazilian Association of 
technical standard: ABNT NBR IEC 60079-1, Table 3.
The enclosure for the explosion proof transmitter and intrinsically safe barrier must be 
remote mounted (refer to 
if the operating temperature exceeds 1
40 
o
F (60 
o
C)
 (refer to 
)
Cable length (refer to 
EXPLOSION OR FIRE HAZARD
Conduit runs must have a sealing fitting within 18 inches (457 mm) of the enclosure to reduce the risk of 
an explosion or a fire. 
During operation, keep covers tight. 
During equipment maintenance, disconnect power before opening transmitter or base 
electronics. Clean cover joints before replacing. 
DO NOT substitute meter components.  Component substituting may compromise the intrinsic 
safety.
Failure to do so may result in severe injury to personnel or cause damage to the equipment.