Piping recommendations
Section 2: Mechanical installation
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual 
June 2014
3-9000-759 Rev D
Daniel 3410 Series Ultrasonic Gas Flow Meters should be mounted in horizontal piping 
with the chord paths horizontal  
Normally, the meter body is installed so that the electronics assembly is on the top of 
the meter. If there is insufficient space above the piping for this arrangement, the meter 
can be ordered with extra long transducer cables for remote mounting or the meter 
housing can be installed with the electronics assembly on the bottom.
The mating piping should include temperature measurement connections located a 
minimum of three nominal pipe diameters length down stream of the meter, or per 
AGA Report No. 9.
FAULTY METER INSTALLATION
Correctly install the equipment.
If meter bodies are mounted or oriented differently than specified above, debris or gas may
collect in the transducer ports which could adversely affect the transducer signals, or cause
equipment damage.