CPU MODULE LABELING AND LED INDICATORS
Section 3: Electrical installation
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014
3-9000-759 Rev D
Serial connections
Use a serial cable, Daniel P/N 3-2500-401, to connect to a PC running Daniel MeterLink. The 
cable is designed for RS-232 communications which is the serial Port A default configuration 
(
 field wiring diagram,
 
Daniel Drawing DMC - 005324). The DB-9 end of the cable 
plugs directly into the PC running Daniel MeterLink. The three wires on the other end of the 
cable connect to the CPU Module RS-485/RS-232 terminals. The RED wire goes to RX, the 
WHITE wire goes to TX, and the BLACK wire goes to COM (
 for Port A wiring). 
RS-485, 2-wire connection on Port A, uses TX+ and TX- on the CPU Module and has a ground 
wire. 
When Beldon wire No. 9940 or equivalent is used, the maximum cable length for RS-232 
communications at 9600 bps is 88.3 meters (250 ft.) and the maximum cable length for RS-485 
communication at 57600 bps is 600 meters (1970 ft.).
Port A supports a special override mode which forces the port to use known communication 
values (19200 baud, address 32, RS-232). Note that the protocol is auto-detected. This mode is 
expected to be used during meter commissioning (to establish initial communication) and in 
the event that the user cannot communicate with the meter (possibly due to an inadvertent 
communication configuration change). Alternately, when using Daniel MeterLink with an 
Ethernet port, use Ethernet cable, Daniel P/N 3-3400-079, to connect the PC.   
Table 3-4  Serial Port A parameters
Port/Communication Description
Common features
Port A (Standard)
RS-232
RS-485 Half Duplex
RS-485 Full Duplex
RS-485
1
 
(
2-wire
 communication on 
Port A.)
1.  
RS-485 2-wire connections use TX+ and TX - on the CPU Module
Typically used for general communications with 
a flow computer, RTU (Modbus slave) and 
radios.
RS-485 
— 
2-wire (Half Duplex) connected to 
TX
+
 and TX
 -
Special override mode to force port configura-
tion to known settings.
Supports RTS/CTS handshaking with software-
configurable RTS on/off delay times.
Factory default is RS-232, Address 32, 19200 
baud.
Communications via Daniel MeterLink 
using RS-232 or RS-485 Full Duplex
Software configurable Modbus 
Address (1-247)
Auto-detects TCP/IP and ASCII or RTU 
Protocol
ASCII Protocol:
Start Bits = 1, Data Bits = 7
 
2
Parity: odd or even 1, Stop Bits = 1
Baud Rates: 1200, 2400, 9600, 
19200, 38400, 57600, 115000 bps
RTU Protocol:
Start Bits = 1, Data Bits = 8
Parity: none, Stop Bits = 1
Baud Rates: 1200, 2400, 9600, 
19200, 38400, 57600, 115000 bps
Ethernet
Preferred port for diagnostic communication
via Daniel MeterLink
10 Mbps/100 Mbps
Modbus TCP/IP, Modbus TCP