Emerson 3414 Manuel D’Utilisation

Page de 144
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
Section 3: Electrical installation
3-9000-759 Rev D
June 2014
Frequency/Digital outputs
 59
3.6.1
Frequency/Digital outputs
The meter has three user-configurable selections for configuring either a Frequency output or 
Digital output (FODO). 
FODO1 (four possible parameter configurations)
FODO2 (eight possible  parameter configurations)
FODO3 (eight possible  parameter configurations)
Frequency or Digital Outputs (FODO 1) source
FO1A, DO1A, FO1B, DO1B
Frequency output 1A is based on frequency content  (Actual - Uncorrected Flow Rate)
Frequency output 1B is based on frequency content and Frequency 1 B Phase
Digital output 1A is based on Digital output1A content (Frequency1A Validity or Flow 
Direction)
Frequency or Digital Outputs (FODO 2) source
FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B
Frequency output 1A is based on frequency content  (Actual - Uncorrected Flow Rate)
Frequency output 1B is based on frequency content and Frequency 1B Phase
Frequency output 2A is based on frequency content  (Actual - Uncorrected Flow Rate)
Frequency output 2B is based on frequency content and Frequency 2B Phase
Digital output 1A is based on Digital output1A content (Frequency 1Validity and Flow 
Direction)
Digital output 1B is based on Digital output1B content (Frequency 1 Validity and Flow 
Direction)
Digital output 2A is based on Digital output 2A content (Frequency 2 Validity and Flow 
Direction)
Digital output 2B is based on Digital output 2B content (Frequency 2 Validity and Flow 
Direction)