SECTION 6—OPERATING INSTRUCTIONS
Invacare XPO
2
32
Using the XPO
2
 Portable Concentrator
1. Turn the XPO
2
 On by pressing the On/Off button.
2. Breathe normally through your nose. Breathing through 
your mouth may result in less effective oxygen therapy. 
The XPO
2
 will put out a measured pulse of oxygen each 
time it detects inhalation. The blue flow setting indicator 
light will blink each time an inhalation is detected and the 
XPO
2
 outputs a pulse of oxygen. 
3. Turn the XPO
2
 Off by pressing and holding the power 
On/Off button for approximately two seconds. 
NOTE: The time required for the XPO
2
 to reach maximum oxygen 
concentration after being turned On is approximately 5 minutes. 
FLOW SETTING
AVERAGE PULSE VOLUME (AT 20 
BREATHS PER MINUTE)
1
15 ml
2
23 ml
3
31 ml
4
37 ml
5
42 ml