MODEL DO-03/04
SECTION 5.0
SOFTWARE CONFIGURATION
TABLE 5-1. Program Settings List
Continued on the following page
ITEM
CHOICES
FACTORY SETTINGS
SETPOINTS
A. Alarms (Section 5.1)
1. Alarm 1 (low action)
a. if oxygen (ppm)
-99 to 99 ppm
0 ppm
b. if oxygen (% saturation)
0 to 200%
0 %
c. if temperature
-5 to 130°C
0.1°C 
2. Alarm 2 (high action)
a. if oxygen (ppm)
-99 to 99 ppm
20 ppm
b. if oxygen (% saturation)
0 to 200%
200 %
c. if temperature
-5 to 130°C
130°C
3. Alarm 3 
See alarm 2
See alarm 2
B. Outputs (Section 5.2 and 5.3)
1. Output 1 or 2: 4 mA setting
a. if oxygen (ppm)
-99 to 99 ppm
0 ppm
b. if oxygen (% saturation)
0 to 200%
0 %
c.  if temperature
-5 to 130°C
0.1°C
2. Output 1 or 2: 20 mA setting
a. if oxygen (ppm)
-99 to 99 ppm
20 ppm
b. if oxygen (% saturation)
0 to 200%
200%
c. if temperature
-5 to 130°C
130°C
CONFIGURE
A. Display options (Section 5.4)
1. Measurement
Oxygen, ozone, free chlorine, total chlorine
Oxygen
2. Sensor (Oxygen only)
Rosemount standard, Rosemount biopharm, or 
Rosemount standard
other steam sterilizable
3. Units (Oxygen only)
ppm, ppb, % saturation
ppm
4. Temperature units
°C or °F
°C
5. Output 1
mA or % of full scale
mA
6. Output 2
mA or % of full scale
mA
7. Language
English, Français, Español, Deutsch, Italiano 
English
8. Main display left
See section 5.5
Sensor current
9. Main display right
See section 5.5
Output 1 current 
10 Display contrast
00-99 (darkest)
50
11. Test timeout
On or off
On
12. Timeout value
1 to 60 min
10 min
Table 5-1 lists the default settings and the range of choices available for each setting. To reduce the chance of