MODEL DO-03/04
SECTION 5.0
SOFTWARE CONFIGURATION
5.9 BAROMETRIC PRESSURE
1.  Barometric pressure is used during air calibration. 
Noise rejection 
Barometric pressure
Main sensor cal
Exit
Enter
2.  Press any key to enter the main menu. Move the cursor to "Program"
and press Enter (F4). 
3.
Move the cursor to "Configure" and press Enter (F4).
4.
Move the cursor to "Barometric pressure" and press Enter (F4).
5. Use 
the 
 and  keys to move through the list of items. To make a
change, press Edit (F4). Use the arrow keys to change settings to the
desired value and press Save (F4). 
Bar meas: The analyzer has an on-board sensor that automatically
measures barometric pressure. To bypass the pressure sensor,
select "Manual'. Be sure to enter the desired pressure in the third line.
During air calibration, the analyzer will use the barometric pressure
entered in this step no matter what the true pressure is. 
Bar units: Select the units in which barometric pressure will be dis-
played. 