Technical specifications
Humidifier performance
The following settings have been tested at 22°C, 40% RH ambient conditions: 
Note: The manufacturer reserves the right to change these specifications without notice.
Maximum heater plate temperature
85°C
Temperature cut-out
93°C
Maximum gas temperature
41°C
Dimensions (H x W x D)
Docking station and water chamber: 130 mm x 170 mm x 212 mm 
Weight
Docking station and unfilled water chamber: 786 g
Water capacity
To maximum fill line: 390 mL
Power supply
100-240V 50/60Hz, 0.7-1.6A; 110V 400Hz, 0.8A 
Instantaneous power consumption: <400W 
Average power: <115W
Docking station 
Flame retardant engineering thermoplastic, plated cast aluminium
Water chamber
Injection molded plastic, stainless steel and silicon seal
Operating temperature
+5°C to +35°C
Operating humidity
10–95% non-condensing
Operating altitude
Please refer to your CPAP, bilevel or non-invasive ventilation device 
user guide
Storage and transport temperature
 -20°C to +60°C
Storage and transport humidity
10–95% non-condensing
IEC 60601-1 classification
Class II (double insulation), Type BF Applied Part
Electromagnetic compatibility
Please refer to your flow generator user's manual for details
 CPAP mask pressure
 cm H
2
O
                RH output %
Setting 3                  Setting 6
Nominal system output AH
*
, BTPS
Setting 3                  Setting 6
 
4
>95
>95
16
20
10
>85
>95
14
18
20
>75
>85
13
16
*
AH- Absolute Humidity in mg/L.
BTPS- Body Temperature Pressure Saturated.
248672 H4i User Guide Eur1.book  Page 10  Friday, February 1, 2013  10:57 AM