ResMed 368870/1 2012-10 Manuel D’Utilisation

Page de 13
English
4
Mask and headgear are not positioned 
correctly.
Adjust position of mask and headgear.
Incorrect air tubing selected.
If you are using the SlimLine or Standard air tubing ensure 
that you have the correct air tubing selected via the menu. 
Pressure rises inappropriately
Talking, coughing or breathing in an 
unusual manner.
Avoid talking with a nasal mask on, and breathe as normally 
as possible.
Mask cushion is buzzing against the 
skin. 
Adjust the headgear.
Displays error message: 
High temperature fault, refer to user manual
Device has been left in a hot 
environment.
Allow to cool before re-use. Disconnect the power cord 
and then reconnect it to restart the device.
Air filter is blocked.
Replace your air filter. Disconnect the power cord and then 
reconnect it to restart the device.
Air tubing is blocked.
Check your air tubing and remove any blockages. 
Disconnect the power cord and then reconnect it to restart 
the device.
Humidifier setting is too high, resulting 
in accumulation of water in the air 
tubing.
Turn the humidifier setting down and empty the water from 
the air tubing.
Displays error message: 
Check ResMed 30/90W Power Supply Unit and fully insert the 
connector
The DC plug is partially inserted into the 
back of the device or inserted too 
slowly.
Fully insert the DC plug.
A non-ResMed power supply unit is 
connected to the device.
Remove the power supply unit and replace with a ResMed 
power supply unit.
The power supply unit is being covered 
by bedding.
Make sure that the power supply unit is free from bedding, 
clothes or other objects that could cover it.
Displays error message: 
Tube blocked, please check your tube
Air tubing is blocked.
Check your air tubing and remove any blockages.
Problem/Possible cause
Solution