ResMed Stellar 150 Manuel D’Utilisation

Page de 40
27
Technical specifications
En
glis
h
Technical specifications
FiO
2
 sensor is used or defective.
If the lifetime of the FiO
2
 sensor has 
exceeded one year please replace the 
FiO
2
 sensor and start calibration again.
LCD: 
 is displayed in the header.
Battery is not charging.
• Ensure the ambient temperature 
conditions are within the specified 
operating range. If the problem persists 
within the specified operating conditions 
please return the device for servicing.
• Power off the device. Power son the 
device again.
Operating pressure 
range
• IPAP: 2 cm H
2
O to 40 cm H
2
O (in S, ST, T, PAC mode) 
• PS: 0 cm H
2
O to 38 cm H
2
O (in S, ST, T, PAC mode) 
• EPAP: 2 cm H
2
O to 25 cm H
2
O (in S, ST, T, iVAPS, PAC 
mode) 
• CPAP: 4 cm H
2
O to 20 cm H
2
O (in CPAP mode only) 
• Min PS: 0 cm H
2
O to 20 cm H
2
O (in iVAPS mode) 
• Max PS: 0 cm H
2
O to 30 cm H
2
O (in iVAPS mode)
 
Note: iVAPS mode is only available in Stellar 150.
Maximum single fault 
pressure
60 cm H
2
O (in all modes) 
Maximum breathing 
resistance under single 
fault 
2 cm H
2
O at 30 L/min; 
7.2 cm H
2
O at 60 L/min 
Maximum flow
> 200 L/min at 20 cm H
2
O  
Flow accuracy
± 5 L/min or 20% measured value, whichever is greater
Test condition: T mode, IPAP: 40 cm H
2
O, EPAP: 2 cm 
H
2
O, Rise Time: MIN, Fall Time: MIN, Ti: 4.0 sec, 
Respiratory Rate: 10 bpm, with ResMed calibration cap. 
Therapy pressure 
tolerance
IPAP: ± 0.5 cm H
2
O ± 10% of set pressure (end of 
inspiration)
EPAP/PEEP: ± 0.5 cm H
2
O ± 4% of set pressure
CPAP: ± 0.5 cm H
2
O ± 10% of set pressure
 
Test condition: T mode, IPAP: 40 cm H
2
O, EPAP: 2 cm 
H
2
O, Rise Time:  MIN, Fall Time: MIN, Ti: 4.0 sec, 
Respiratory Rate: 10 bpm, with ResMed calibration cap. 
Sound pressure level
29 dBA as measured according to ISO 17510 – 1; 2002.
32 dBA with uncertainty of 3 dBA as measured according 
to ISO 17510 – 1:2007.
Alarm Volume Range
> 45 dBA - <85 dBA at 1 meter (3 steps: low, medium, high) 
Dimensions (L x W x H)
230 mm x 170 mm x 120 mm
Weight
2.1 kg
Air outlet
22 mm taper, compatible with ISO 5356-1:2004 
Anaesthetic & Respiratory Equipment - Conical 
Connectors
Pressure measurement
Internally mounted pressure transducer
Flow measurement
Internally mounted flow transducer
Problem / possible cause
Solution