ResMed Stellar 150 Manuel D’Utilisation

Page de 40
5
Setting up for invasive use
En
glis
h
Note: Other external humidifiers can be used for noninvasive setup. ResMed 
recommends using a breathing system (including antibacterial filter, air tubings, external 
humidifier) with an impedance of maximum 2 cm H
2
O at 30 L/min, 5 cm H
2
O at 60 L/min 
and 16 cm H
2
O at 120 L/min.
4
Connect one end of the air tubing firmly onto the air outlet of the H4i.
5
Connect the mask system to the free end of the air tubing. 
6
Select the mask type (select Setup menu, then Clinical Settings, then Advanced Settings).
7
Attaching the H4i heated humidifier for noninvasive use
Use of the H4i heated humidifier may be recommended by your clinician. For information 
on using the H4i, see the H4i user guide.
WARNING
• Always place the H4i on a level surface below the level of the patient to prevent 
the mask and tubing from filling with water.
• Make sure that the water chamber is empty and thoroughly dried before 
transporting the humidifier.
• The H4i heated humidifier is not intended for mobile use.
• Do not overfill the water chamber, as during use this will cause water to spill into 
the air circuit.
• For optimal accuracy and synchrony, perform Learn Circuit with a change of the 
circuit configuration, in particular when adding or removing high impedance 
components (eg, antibacterial filter, external humidifier, water trap, nasal pillow 
type mask or air tubing). See “Setup menu: Options” on page 16.
CAUTION
Check the air circuit for water condensation. Use a water trap or a tubing wrap if  
humidification is causing water condensation within the tube. 
Notes:
• A humidifier increases resistance in the air circuit and may affect triggering and cycling, and 
accuracy of display and delivered pressures. Therefore perform the Learn Circuit function 
(see “Setup menu: Options” on page 16).
 The device adjusts the airflow resistance.
• The heating feature of the H4i is disabled when the device is not mains powered.
Setting up for invasive use
The Stellar can be used invasively only with the ResMed Leak Valve.
WARNING
• Make sure that all the air inlets at the rear of the device and under the device and 
vents at the mask or at the ResMed Leak Valve are unobstructed. If you put the 
device on the floor, make sure the area is free from dust and clear of bedding, 
clothes or other objects that could block the air inlets.
• When using a Heated Moisture Exchange Filter (HMEF), replace the HMEF 
regularly as specified in the instructions provided with the HMEF.
• The H4i is contraindicated for invasive use. An external humidifier approved for 
invasive use is recommended according to EN ISO 8185 with an absolute 
humidity of > 33 mg/L.
• For optimal accuracy and synchrony, perform Learn Circuit with a change of the 
circuit configuration, in particular when adding or removing high impedance 
components (eg, antibacterial filter, external humidifier, water trap, nasal pillow 
type mask or air tubing). See “Setup menu: Options” on page 16.