Guidance and manufacturer’s declaration – electromagnetic immunity
The VPAP device is intended for use in the electromagnetic environment specified below. The customer or the user of the VPAP 
device should assure that the device is used in such an environment.
Immunity test
IEC60601-1-2 test 
level
Compliance level
Electromagnetic environment – guidance
Electrostatic 
discharge (ESD)
IEC 61000-4-2
±6 kV contact 
±8 kV air
±6 kV contact 
±8 kV air
Floors should be wood, concrete or ceramic tile. If 
floors are covered with synthetic material, the relative 
humidity should be at least 30%.
Electrical fast 
transient/burst
IEC 61000-4-4
±2 kV for power 
supply lines
±1 kV for input/
output lines
±2 kV 
Not Applicable
Mains power quality should be that of a typical 
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV differential 
mode
±2 kV common 
mode
±1 kV differential 
mode 
±2 kV common 
mode
Mains power quality should be that of a typical 
commercial or hospital environment.
Voltage dips, short 
interruptions and 
voltage variations 
on power supply 
input lines. 
IEC 61000-4-11
<5% Ut (>95% dip in 
Ut) for 0.5 cycle 
40% Ut (60% dip in 
Ut) for 5 cycles 
70% Ut (30% dip in 
Ut) for 25 cycles 
<5% Ut (>95%dip in 
Ut) for 5 sec 
< 12V (>95% dip in 
240V) for 0.5 cycle 
96V (60% dip in 
240V) for 5 cycles 
168V (30% dip in 
240V) for 25 cycles 
<12V (>95%dip in 
240V) for 5 sec
Mains power quality should be that of a typical 
commercial or hospital environment.
If the user of the VPAP device requires continued 
operation during power mains interruptions, it is 
recommended that the VPAP device be powered from 
an uninterruptible power source.
Power frequency 
(50/60Hz) magnetic 
field IEC 61000-4-8
3 A/m 
3 A/m 
Power frequency magnetic fields should be at levels 
characteristic of a typical location in a typical 
commercial or hospital environment.
Conducted RF 
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 
150 kHz to 80 MHz
10 V/m
80 MHz to 2.5 GHz
3 Vrms 
10 V/m 
Portable and mobile RF communications equipment 
should be used no closer to any part of the VPAP 
device, including cables, than the recommended 
separation distance calculated from the equation 
applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.17 
√P
d = 0.35 
√P 80 MHz to 800 MHz
d = 0.70 
√P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer and d is the recommended separation 
distance in meters (m).
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey,
a
 
should be less than the compliance level in each 
frequency range.
b
 
Interference may occur in the vicinity of equipment 
marked with this symbol 
.
NOTE 1: Ut is the AC mains voltage prior to application of the test level.
NOTE 2: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 3: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 
from structures, objects and people.
a
 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, 
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the 
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured 
field strength in the location in which the VPAP device is used exceeds the applicable RF compliance level above, the VPAP device 
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such 
as reorienting or relocating the VPAP device.
b
 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.