Cleaning and maintenance
En
glis
h
.
Disinfection
Disinfection of your device helps to prevent the risk of cross contamination.
Disinfect the exterior of the device, and especially the air outlet, with a damp cloth and a 
disinfection solution (eg, Microzid).
Multipatient use
WARNING
• An antibacterial filter is mandatory if the device is used on multiple patients.
• In a mulitpatient use environment, you must perform the following before the 
device is provided to a new patient:
Air filter cover
Air filter
Air filter and the 
antibacterial filter
Replace.
Mask
Reprocess; Cleaning, disinfection and sterilisation 
instructions are available from the ResMed website, 
www.resmed.com/masks/sterilization. If you do not have 
Internet access, please contact your ResMed representative.
Air tubing
Replace the air tubing. Alternatively, consult the air tubing 
instructions for cleaning and disinfection information.
Device
Disinfect the Stellar as follows:
Use an anti-bacterial cleaning/disinfectant solution such as 
Mikrozid
®
 AF or CaviCide
®
 with a clean non-dyed 
disposable cloth to clean and disinfect the exterior surfaces 
of the device. Wipe all accessible surfaces of the device, 
including the air outlet (avoid liquid entering any openings in 
the device). Follow the manufacturer’s recommended 
cleaning instructions.
Humidifier
As instructions for humidifiers vary, see the user guide for 
the humidifier in use. For multipatient use environment, use 
the H4i water chamber (disposable) in place of the H4i 
reusable water chamber.