USING THE SCALE
23
Calibrating the 1100 Bath Scale
NOTE: For this procedure, refer to FIGURE 3.1 on page 20 and FIGURE 3.4.
1. The patient MUST be removed from the bath lift to properly calibrate the bath scale. Refer 
to Transferring into/out of the Lift of the Bath Lift Owner’s Manual for instructions.
2. Remove the seat/stretcher from the bath lift.
• Lock the two rear locking casters.
• Press on the down arrow on the hand pendant to lower the bath lift.
• Remove the three bolts that secure the hanger rods to the bath scale.
• Pull the hanger rods away from the patient scale to remove the seat or stretcher from 
the bath lift.
3. Remove the four screws that attach the cover to the bath scale.
NOTE: The RUN/CAL switch is located next to the display window and is visible after the cover is 
removed.
4. Slide the switch marked RUN/CAL to the CAL position.
5. Adjust R9 so the display indicates +15 pound offset. Refer to Detail “A” of FIGURE 3.4.
6. Reverse STEP 2 to attach the seat or stretcher to the bath lift.
7. Make note of the weight of the seat or stretcher on the display.
8. Put 200 pounds of calibrated weight on the seat or stretcher. Check for the proper weight 
increase relative to the weight of the seat or stretcher noted in STEP 7.
9. Adjust R18 until the weight on the display is accurate of the combined weight of the 
calibrated weight and the weight of the seat or stretcher. Refer to Detail “A” of FIGURE 3.4.
10. Remove the calibrated weight from the bath scale. Repeat STEPS 5‐9 as necessary.
11. Slide the RUN/CAL switch to the RUN position. Press the ZERO button.
12. When the display window reads 0.0, place the calibrated weight on the seat or stretcher 
and make sure the weight on the display is correct.
13. Remove the calibrated weight from the bath scale.
NOTE: The bath scale is ready for use.
FIGURE 3.4   Calibrating the 1100 Bath Scale
Screw
(1 of 4)
Cover
R9 Adjustment Screw
R18 Adjustment Screw
DETAIL “A” - 1100 BATH SCALE CIRCUIT BOARD