Invacare 1150703 Manuel D’Utilisation
SECTION 7—MAINTENANCE
25
NOTE: The actuator lead wires will be visible after the large base cover is removed.
6.
Remove the socket head screw and washer that attach the leg actuator to the base.
7.
At the other end of the actuator, remove the socket head screw and nut that attach the leg actuator to the leg bracket.
8.
If necessary, repeat STEPS 5-7 on the opposite side of the base to remove the other leg actuator.
9.
To replace the actuator(s), reverse STEPS 5-7.
FIGURE 7.2 Replacing the Leg Actuators
Adjusting the Knee Pad Height
NOTE: For this procedure, refer to FIGURE 7.3.
NOTE: When adjusted properly, the patient’s knee will make
contact at the center of the knee pad.
1.
NOTE: When adjusted properly, the patient’s knee will make
contact at the center of the knee pad.
1.
While supporting the knee pad, disengage the quick
release knob on the side of the knee pad mounting
bracket.
release knob on the side of the knee pad mounting
bracket.
2.
Choose a height adjustment that will be comfortable for
the patient and will provide the necessary support.
Engage the quick release knob into position until the knee
pad is secure in place.
the patient and will provide the necessary support.
Engage the quick release knob into position until the knee
pad is secure in place.
Small Base Cover
Mast Handles
BOTTOM VIEW OF
PATIENT LIFT
Large Base Cover
Screw
(eight screws attach each
small base cover)
Pan head screw
(eight screws attach the
large base cover)
Leg
Caster
C
L
OSE
Knee Pad
Quick Release
Knob