Sigma LBA-400 Manuel D’Utilisation

Page de 281
 
 
Figure 19 
 
 
Operating hints: 
•  Increasing Video Gain also increases video noise.  Use Video Gain sparingly, or not at all. Leave 
it set to 1 whenever possible. 
•  The Black Level will be adjusted automatically each time you perform an Ultracal! calibration 
cycle.  Therefore, it is best that you make no changes to this setting.  If you alter the Black Level 
after an Ultracal!, the calibration will no longer be valid. 
 
3.2.4.3.1 
The Four Camera Option 
Some enhancements have been made to the four-camera option with the release of software 
version 4.00.  These enhancements make it possible to automatically sequence data 
collection between multiple cameras (up to four).  This sequencing only applies to analog 
cameras, and the following operating conditions must be met: 
•  All cameras must be the same type or model. 
•  The cameras must be Gen-Locked together. 
•  If cameras with remotely controlled shutters are used the shutter control circuits 
must be paralleled to each camera. 
The status of the input video is shown in the video enunciator located at the bottom of the 
LBA-PC’s main display window.  The number shown in this enunciator indicates which camera 
is currently selected.  If it shows 
GREEN
 then camera sync is present.  If it shows 
RED
 then 
there is either no camera connected, or the connected camera is not operating. 
3.2.4.3.2 
How to Specify Video Gain, Black Level, Shutter, and Video Trigger Level 
On the right side of the Camera panel you will see a panel containing a Camera# drop-
down, Video GainBlack Level, and Shutter controls.  Click on the drop-down to select 
which camera is affected, the current settings are displayed, and then make the desired 
changes.  Note that Video Trigger Level, Trigger section below, can also be set for each 
camera.   
On the toolbar, changing the Video GainBlack LevelShutter, and Video Trigger Level 
affects only the current camera listed in the video enunciator located at the bottom of the 
main display window. 
Operator’s Manual 
 
LBA-PC 
 
Doc. No. 10654-001, Rev 4.10 
61