Sensor Installation: F-Series
Mounting the Sensor 
Mounting the electronics of high-temperature and extreme high-temperature sensors
High-temperature and extreme high-temperature F-Series sensors come with a 32
″ (812 mm) pre-installed 
flexible conduit. This conduit is required for agency approval. 
A factory-supplied Model 1700/2700 transmitter, core processor, or junction box is connected to the end of 
the flexible conduit. Mount the electronics to a wall or instrument pole using the supplied bracket (Figure 10).
Figure 10
Electronics mounting for high-temperature F-Series sensors 
Mounting 
bracket 
(wall mount)
Use 4 × 5/16
″ (8 mm) 
bolts to secure the 
bracket to a wall
Slide the electronics into the
bracket groove.
Mounting 
bracket
(instrument pole)
Use 2 × 5/16
″ (8 mm) 
U-bolts to secure the 
bracket to an 
instrument pole
Secure the electronics to the 
bracket with the 4 #10 screws.