Emerson NGA2000 PMD Manuel D’Utilisation

Page de 88
Instruction Manual 
760007-A 
July 2003 
Rosemount Analytical Inc.    A Division of Emerson Process Management                                                                                 Installation     2-7 
Model NGA2000 PMD 
.6
[15]
1.6
[40]
20.0
[508]
17.41
[442]
8.4
[213]
6.7
[170]
1.6
[41]
1.0
[25]
IN
OU
SAMPLE
10 PSI MAX
(69 kP
2.8
[71]
.5
[13]
4.3
[109]
1.1
[28]
6.0
[152]
.6
[15]
8.2
[208]
8.4
[213]
6.2
[157]
DIMENSIONS 
 
INCH 
[mm] 
E.  FUSE. 
D.  POWER CABLE TO NETWORK. 
C.  NETWORK CABLE CONNECTIONS TO PLATFORM. 
B.  SAMPLE OUT:  1/4" O.D. TUBE FITTING. 
A.  SAMPLE IN:  1/4" O.D. TUBE FITTING. 
 
5.  MODULE TO BE INSTALLED WITHIN 
±15° OF HORIZONTAL. 
4.  POWER REQUIREMENTS: 24 VDC  3.5 A. 
3.  ELECTRICAL INSTALLATION MUST BE IN COMPLIANCE WITH NATIONAL ELECTRICAL 
     CODE (ANSI/NFPA 70) AND/OR ANY APPLICABLE NATIONAL OR LOCAL CODES. 
2.  MODULE IS NOT WEATHERPROOF. 
1.  APPROXIMATE WEIGHT: 15 LB (6.8 kg). 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 2-6.  PMD Outline and Mounting Dimensions