Specifications
Audible/Visual Alarm Status Indicators: Four Moisture levels, Temperature, 
Moisture cell Diagnostic, Zero Verification or Calibration in Process, Moisture Analyzer 
off-line, Analog output freeze during calibration.
Alarm Relays: Four non-latching, independently assignable to moisture alarms or to 
moisture calibration-in-process indicator. SPDT contacts rated at 1 Amp @ 30 VDC.  Fail 
safe action upon loss of power to alarm condition. Not designed to switch AC power.
Storage Temperature: Not to exceed 50° C (122° F)
Construction: NEMA 1, 19 inch rack mount
Dimensions: 19 inch (48.3cm) wide x 10.5 inch (26.7 cm) high x 22.5 inch (57.2 cm) 
deep
Weight: 70 pounds (31.8 kg)
Gas Path Construction Materials:
300 series stainless steel electro-polished
1/4-inch 
VCR
-type compatible sample inlet fitting
1/8-inch compression sample outlet fitting
1/4-inch compression vacuum fitting
Kel-f valve seats