Installation, Start Up and Shut Down                                   DF-745
19
4.2 Analyzer Start Up
It is important to note that, if equipped with automatic gas valves, the moisture cell is 
isolated from gas flow while the analyzer is off power.
Open the door and turn on the power using the main power switch inside the analyzer. See 
Figure 2. The pump, if equipped, should turn on (see page 17) and the analyzer will 
undergo a series of Diagnostic Procedures while the various startup screens are displayed. 
Next, the Delta F Corporation logo is briefly displayed and then the data display appears 
with the “Warming Up” screen flashing. The warm up process takes approximately six to 
ninety minutes after which the display will look similar to Figure 8 (values shown are only 
representative).
During the Warm Up process the gas valves (if equipped) can be opened and gas flow 
started. See the section on Gas Pressure and Flow Settings on page 20 for important 
information on the gas delivery system and setting proper pressures and flow rates.
NOTE
During the warm up period all analog and digital outputs are held to an 
artificial 0.011ppb reading to avoid the reporting of false readings.
Figure 8: Data Display Screen
4.2.1 Gas Delivery System
The gas delivery system as shown in Figure 9 is designed to deliver a gas flow rate of 2 
liters per minute to the moisture cell while maintaining the highest standards of gas purity 
and delivery for ultra-trace analysis.  Features include a single inlet line and flow meter for 
the gas sample, and optional bypass loop to maintain constant purging, and essentially dead-
leg free delivery.  
The connections at the rear of the gas delivery system include a ¼” VCR swivel connection 
for the process inlet, a ¼” compression outlet for the moisture cell, an optional pneumatic 
gas connection (1/8” compression), and an optional ¼” compression outlet for the bypass 
loop.  Also on the rear of the gas delivery system, if equipped, are a sample gas regulator to 
adjust the internal sample pressure and a bypass flow meter.