Appendix B – Hydrogen Service Safety System
CAUTION
The operator is obligated to assure proper operation of the analyzer 
air flow system as designed. Do not impede air flow at the inlet in 
the front door or at the exhaust fan outlets on either side of the 
cabinet in the rear.
NOTES
For hydrogen, which has a lower explosive limit (LEL) of 4%, the maximum allowable 
influx in event of an internal leak would be 120 scfh, whereas the normal flow as defined in 
the Operating Instruction Manual is about 1/10
th
that at approximately 14 scfh.
2
Use of the analyzer case purge is at the customer’s discretion.  It most likely is not 
necessary since in the event of sample delivery interlock, the sample feed is blocked from 
entering the chassis and the only open flow path through the system is dissipation of 
remaining sample in the bypass, which would be vented externally, or in the event of an 
internal leak, would be volumetrically insignificant.
3
The pneumatic control for the sample interlock and case purge valves is provided by an 
internal 12 VDC solenoid.