Bacharach INSIGHT Manuel D’Utilisation

Page de 31
Instruction 24-9460
16
Fyrite
®
 INSIGHT
Instruction 24-9460
17
Fyrite
®
 INSIGHT
Change the order in which data is displayed as follows:
1. Enter the SETUP MENU per Section 3.5.1.  If necessary, press ESC 
until MENU appears above F2.
2. Use the ▲▼ buttons to select RUN/HOLD 
FORMAT.  Press ENTER to display the 
Run/Hold Format options.
3. Use the ▲▼ buttons to select EDIT FOR-
MAT.  Press ENTER to display the current 
format.
4.  Change data for a particular location by 
fi rst using the ▲▼ buttons to select the loca-
tion.  Press ENTER and the cursor will start 
to fl ash.
5. Use the ▲▼ buttons to scroll through and 
select the desired data to appear in that 
location.  Press ENTER to save the selection.
6. Change the data displayed at other locations 
by repeating steps 4 and 5
7.  When fi nished, use the ▲▼ buttons to select EDIT COMPLETE, 
located at the bottom of the list.  Press ENTER to save the new dis-
play format and return to the Run/Hold Format options.
Reset Display format back to factory default settings as follows:
1. Enter the SETUP MENU per Section 3.5.1.  
If necessary, press ESC until MENU ap-
pears above F2.
2. Use the ▲▼ buttons to select RESET FOR-
MAT.  Press ENTER to display the Reset 
Format option.
3. Use the ▲▼ buttons to select YES.  Press 
ENTER
 to reset the format and return to the 
Run/Hold format options.
3.5.10 Language Selection
Information on the display screen can be shown in English, French, or 
Spanish.  
Select the desired language as follows:
1. Enter the SETUP MENU per Section 3.5.1.  If necessary, press ESC 
until MENU appears above F2.
2. Use the ▲▼ buttons to select LANGUAGE 
SELECTION.  Press ENTER to display the 
Language Selection Menu.
3. Use the ▲▼ buttons to select the desired 
language.  Press ENTER to activate the 
selection and return to the SETUP MENU.
3.5.11 Button Sound
The audible sound used to signal when a button is pressed can be turned 
OFF and ON as follows:
1. Enter the SETUP MENU per Section 3.5.1.  If necessary, press ESC 
until MENU appears above F2.
2. Use the ▲▼ buttons to select BUTTON 
SOUND.  Press ENTER to display the Button 
Sound Menu.
3. Use the ▲▼ buttons to select YES or NO 
button sound.
4. Press ENTER to save the selection and re-
turn to the SETUP MENU or ESC to revert 
to the previous setting.
3.5.12 CAL Reminder Period
The analyzer can be set to indicate a calibration reminder during the 60 
second warm-up.  Calibration reminders can be preset to occur never, 6, 
8, 10, 12, or 15 months after the last calibration.  When the preset period 
is exceeded the instrument will display the reminder, and how long since 
the sensors were last calibrated.  If a calibration reminder is displayed the 
operator can press the RUN/HOLD key to move to the Run/Hold Screen for 
normal operation.  Regular calibration periods of 6 months to 1 year are 
recommended.