Bacharach INSIGHT Manuel D’Utilisation

Page de 31
Instruction 24-9460
4
Fyrite
®
 INSIGHT
Instruction 24-9460
5
Fyrite
®
 INSIGHT
1.3 Operational Overview
Pressing the PWR button turns the analyzer ON. Note that there is a 
warm-up period of 60 seconds that must elapse before the analyzer can be 
used.
To perform a combustion test, choose a fuel code that corresponds to the 
fuel being burned by the appliance being tested (Section 4.3), and then 
press the RUN/HOLD button to place the analyzer into its Run Mode. 
 Begin testing by inserting the analyzer’s probe tube into the flue-gas 
stream of the appliance under test. The analyzer will monitor the flue 
gas and display measured and calculated values that are relative to the 
combustion process. These values are dis played on the analyzer’s LCD, and 
are chosen for display by pressing the In cre ment () and Dec re ment () 
but tons.
Pressing the RUN/HOLD button during a test will freeze all measured and 
calculated values at their current levels. Pressing the RUN/HOLD button 
again resumes testing.
A backlight enables an operator to read the display in dimly-lit areas. 
Pressing the PWR button turns the backlight ON and OFF. 
An optional power saver function will shut the analyzer OFF after a set 
time period of inactivity. This feature is dis abled if the CO value is greater 
than 50 ppm or the O
2
 value is less than 16.0%.
Press the PWR button for at least two seconds to power the analyzer 
OFF. Note that there is a 5 second delay before the analyzer actually 
turns OFF, during which time the unit can be kept ON by pressing the 
RUN/HOLD 
button. There is a CO purge feature that keeps the analyzer’s 
pump running until the measured CO level drops below 50 ppm.
2.0 TECHNICAL CHARACTERISTICS
The Analyzer measures & displays:
Primary/Ambient Air Temperature
  -4 to 600 °F (-20 to 316°C)
Oxygen
0.1 to 20.9% O
2
Stack Temperature
-4 to 1202 °F (–20 to 650 °C)
Carbon Monoxide
0 to 2000 ppm CO
Differential Pressure/Draft
± 100 inwc
The Analyzer computes & displays:
Combustion Efficiency
0.1 to 100%
Carbon Dioxide
0.1 to a fuel dependent  
maximum value in %
CO referenced to %O
2
0 to 9,999 ppm
Excess Air
0 to 250%
Fuel Selection
• Natural Gas
• Oil #2 
• Oil #4
• Oil #6
• Propane
• Kerosene
• Wood
• Coal
Normal Operating Conditions:
Temperature
    • Analyzer
    • Probe
32 to 104 °F (0 to 40 °C)
1,000 °F max (538 °C) at 
5" insertion
Humidity
   15 to 90% RH, Non-Condensing
Dimensions
Weight
7.9 in (20.1 cm) Height      
3.6 in (9.1 cm)  Width
2.3 in (5.8 cm) Depth
1lb. (0.45 kg)