1.  Enter the main menu by activating any key to get the “enter main menu” prompt, then activate switch 1 to 
select “yes”. 
 
2.  Activate the up key (switch 1) or down key (switch 2) until “Relay Assignment?” is displayed.  
 
3.  Activate the enter key (switch 3) to enter the option. The sub menu: ‘Alarm Relay 1’, ‘Alarm Relay 2’, ‘Alarm 
Relay 3’ as well as ‘Exit’ will be displayed. 
 
4.  Choose the Alarm relay (Alarm relay 1, Alarm relay 2, Alarm relay 3) for configuration, by using the up- 
down arrow keys. 
 
5.  Activate the enter key (switch 3) to make the selection. The relay (RL) and channel (CH) with alarm level 
setting will be highlighted. The alarm level settings available are:  Point 1, Point 2, and Disabled.  
 
6.   Under the specific relay with channel, activate the enter key (switch 3) to choose the appropriate setting. See 
Table 3 below. 
 
 
      Table 3
: Available Millennium II Relay Options 
 
 
7.  Once the appropriate settings are chosen, use the up or down arrow key (switch 1) or (switch 2) and switch 3 
to exit.    
 
Example: LEL combustible sensor connected to channel 1 & H2S sensor connected to channel 2. 
Step 1- Set Alarm Levels (points): Setup alarm levels for each channel, e.g. Channel 1 point 1 = 20% LEL, Channel 1 
point 2 = 40% LEL,  Channel 2 point 1 = 10 ppm, and Channel 2 point 2 = 20 ppm. Refer to Section ‘4.2.4: Viewing 
and setting alarm levels (points)’ when setting alarm levels. 
 
Step 2 - Set Relay Options:  Each alarm relay can be set up for energized or de-energized and latching or non-latching. 
Configure alarm relays options as desired. See ‘4.2.5 Setting Relay options’. 
 
Step 3 - Relay Assignment: There are (3) alarm relays and (2) channels. Assign alarm relays to channels as desired, 
e.g. 
  Alarm relay 1 assigned to Channel 1 Point 1 (20% LEL), 
  Alarm relay 2 assigned to Channel 2 Point 1 (10 ppm) & 
  Alarm relay 3 assigned to Channel 1 Point 2 (40% LEL) 
  Alarm relay 3 assigned to Channel 2 Point 2 (20 ppm). 
 
Relays and  Assignment Options 
Channel # and Alarm 
points(levels) 
ALARM RELAY 1 (RL1) 
with channel # and Alarm 
points (levels) 
ALARM RELAY 2 (RL2) 
with channel # and Alarm 
points (levels) 
ALARM RELAY 3 (RL3) 
with channel # and Alarm 
points (levels) 
CH1 
POINT 1= Alarm level 1 
POINT 2= Alarm level 2 
RL1:CH1 
Point 1/Point 2/Disabled 
RL2:CH1 
Point 1/Point 2/Disabled 
RL3:CH1 
Point 1/Point 2/Disabled 
CH2 
POINT 1= Alarm level 1 
POINT 2= Alarm level 2 
RL1:CH2 
Point 1/ Point 2/Disabled 
RL2:CH2 
Point 1/ Point 2/Disabled 
RL3:CH2 
Point 1/Point 2/Disabled