2-9
CCO 5500
POWER SUPPLY UNIT 
Enough cable is supplied to mount the power supply up to a maximum of 33 ft 
(10 m) from the transmitter. A maximum total cable length of 82 ft (25 m) may 
be used to link the power supply to the receiver. These are the maximum 
permitted lengths of cable and must not be exceeded. 
Dimensions and mounting hole locations are identical to the signal processor 
and are illustrated in Figure 2-4. 
ELECTRICAL 
CONNECTIONS 
Installation of Cables 
Decide routing for all non-power cables (both those supplied by Rosemount 
Analytical and those sourced locally). Use common routing wherever possible 
and install leaving sufficient free-end length to make final connections. 
Power cables should be installed separately, using different routes if possible 
to reduce the risk of cross interference. Leave sufficient free-end length to 
make final connections. 
Rosemount supplied cables are provided with ferrite beads fitted to all cores 
to protect against interference and should not be modified without consulting 
Rosemount. 
Cable Connections
Overall system connections are illustrated in Figure 2-5.
Wiring should only be undertaken by a qualified technician.
Ensure that the power supply is isolated. 
DO NOT switch power on until all installation work is complete and the system is ready for 
configuring.